F500 Anaerobic/Hypoxic Workstation

Used for inoculation, continuous cultivation, passage, medium transformation, observation, and identification of microorganisms in anaerobic (microaerophilic) environments. Shigella, Campylobacter jejuni, Clostridium perfringens, Clostridium botulinum, and Lactobacillus can be cultured simultaneously according to national standards.

Description

*1.  Equipped with a built-in quick transfer gate with a capacity of 28L, the internal dimensions of the transfer gate (W * D * H) are 300x335x280mm, and it can transfer 90 90mm petri dishes at once, which can transfer most anaerobic tanks on the market; Transfer gate inner door size: 180x280mm, transfer gate outer door size: 230x330mm; Also equipped with a single dish transfer system.

2.  Equipped with transfer gate intelligent protection function, the interlocking linkage between the inner and outer doors ensures that they cannot be opened simultaneously, ensuring that the internal environment is not affected.

3.  The transfer gate uses N2 for purging, and parameters such as the nitrogen purging time (0-999 seconds), vacuum replacement times (0-99 times), and vacuum degree (0-70%) of the transfer chamber can be freely set on the screen.

4.  Internal dimensions of the operation compartment (W * D * H): 900x530x620mm, capable of accommodating 900 90mm petri dishes; The main body of the operation warehouse is made of food grade safe and non-toxic board with good sealing performance.

*5.  The front panel is made of 20mm thick acrylic panel, with a 10 ° slope that conforms to ergonomic design; It can be lifted upwards as a whole, using dual pressure support rods, easy to open and close, convenient for transferring equipment or a large number of samples into/out of the operation compartment, or for deep cleaning, maintenance or inspection of the equipment. Opening size (width * height): 750mm * 470mm, internal depth of 700mm, used for the entry and exit of small equipment and cleaning of the box.

6. Parameter settings, device status monitoring, and alarms are all implemented on a color 10.1-inch touch screen, with one click operation. One click can complete the opening and closing of oxygen concentration, carbon dioxide concentration, temperature, humidity, pressure, fan, lighting, and other items. Clicking on ‘One click Stop’ on the main interface can close all controls; Having three-level permission password login control, device management is secure and efficient; Display manufacturer information on the touch screen; Chinese and English operating interface.

7.  Equipped with ultrasonic humidification system, fully automatic semiconductor humidity control system, humidity control range: ambient humidity -90%, with low water level automatic stop function.

8.  Equipped with a built-in HEPA (U16) filter, it can meet the ISO14644-1 class 3 standard.

*9.  Oxygen, carbon dioxide, and hydrogen control range: Oxygen control range: anaerobic mode below 5ppm, low oxygen mode 0-100%, accuracy+0.1%; Supports displaying carbon dioxide control range in two forms: percentage and absolute oxygen partial pressure (mmHg): 0-30%, increasing by 0.1%. Optional hydrogen control module, control range 0-10%, increasing by 0.1%, O2, CO2 and H2 sensor comes with a calibration mode, eliminating the need for manual calibration;

10. Equipped with reservation function, it can reserve the on/off of categories such as O2、 CO2, temperature, humidity, lighting, ultraviolet lamps, purification, etc., with a reservation time of 0-20160 minutes and real-time display of the remaining reservation time. Standard configuration: 1 LED lighting fixture with adjustable brightness levels of 5; Two 254nm ultraviolet sterilization lamps can disinfect HEPA and the gas inside the operating chamber. The disinfection time of the ultraviolet lamps can be adjusted independently, with a maximum setting of 20160 min.