D500G Anaerobic/microaerophilic workstation

Used for inoculation, continuous cultivation, passage, medium conversion, observation, identification, and other tasks of microorganisms in anaerobic (microaerophilic) environments. It can simultaneously cultivate microorganisms such as Shigella (anaerobic, 41.5 ℃), Campylobacter jejuni (microaerophilic, 25 ℃), Clostridium perfringens (anaerobic, 36 ℃), Clostridium botulinum (anaerobic, 35 ℃), and lactic acid bacteria (anaerobic, 36 ℃) as specified in the national standard.

Description

 Working conditions:

-Power conditions: 220V, 50/60Hz Current: 4A Environmental conditions: ambient temperature 25 ℃, humidity<90%

 

 Technical parameters:

1. Equipped with a built-in quick transfer gate, with a capacity of 23 liters, it can transfer 60 90mm petri dishes at once; The internal dimensions of the transfer gate (W x H x D) are 290 × 250 × 320mm, and the internal and external door dimensions of the transfer gate are 250 × 250mm. The transfer gate uses N2 blowing to smoothly transfer the sample or equipment out or into the workstation within 60 seconds.

2. The transfer compartment door adopts a unique mechanical press lock to ensure that air does not seep into the box when flushing the transfer compartment.

3. The opening and closing of the inner door of the transfer gate can be operated with one hand and one button, making the operation simple and convenient.

4. Can accommodate 900 90mm petri dishes.

5. The side panel can be removed as a whole for the entry and exit of small devices and the cleaning of the box. Cleaning the box can be done with a specialized workstation disinfectant, which is composed of a mixture of EDTA tetrasodium and 3-tert-butylphenol sodium polyoxyethylene ether sulfate. It can effectively clean and sterilize the box without damaging it.

6. Gas supply: single gas (oxygen free mixed gas, composed of 80% N2+10% H2+10% CO2), or dual gas (one bottle of pure nitrogen and one bottle of oxygen free mixed gas).

7. Parameter settings, device status monitoring, and alarms are all implemented on a color 7-inch touch screen.

8. Equipped with ultrasonic humidification system, fully automatic humidity control system, humidity control range: ambient humidity -90%.

9. Standard fluorescent lamp and 254nm ultraviolet sterilization lamp.

10. Optional built-in HEPA (U16) filter with filtration efficiency up to ISO14644-1 class 3 standard (testing report provided).

11. Equipped with high-precision oxygen indicator, real-time display of oxygen concentration value inside the box.

12. High anaerobic degree, oxygen concentration value below 50ppm (provide actual test pictures, on-site test acceptance).

13. Temperature: Control temperature range: room temperature+3 ℃ to 52 ℃, display accuracy of 0.1 ℃, average temperature ± 0.3 ℃ (provide temperature test report, on-site measurement and acceptance).

14. The long sleeved sleeve can be operated with bare hands. The sleeve system entering the workstation undergoes a process of vacuuming and nitrogen filling, ensuring 100% removal of all oxygen when manually entering the workstation, without any oxygen entering the workstation.

15. Wireless foot switch controls vacuuming and nitrogen filling, with an external vacuum pump as standard.

16. Comes with 2 built-in power sockets as standard.

17. Data recording: A data recording system with a USB interface that can record temperature, humidity, and intake time, with a data recording time of up to 30 days. Provide screenshots of the software interface.

18. The alarm system includes: high and low temperature alarm, humidity range exceeding alarm, anaerobic mixed gas and nitrogen low pressure alarm, gas leakage alarm; Equipped with an automatic gas overcurrent protection device, the intake solenoid valve will automatically close when continuous air intake exceeds 5 minutes. Provide screenshots of the software interface.

19. Timed intake system: equipped with a timed intake system, the intake time can be set arbitrarily and displayed on the touch screen; The fast mode time can be set arbitrarily and displayed on the touch screen. Provide screenshots of the software interface.